Corrosion Resistance Tester adapted to the surface treatment of various materials, including paint, electroplating, organic and inorganic coating, anode processing, antirust oil and so on, after these ...

Inquire Now
Description

Corrosion Resistance Tester adapted to the surface treatment of various materials, including paint, electroplating, organic and inorganic coating, anode processing, antirust oil and so on, after these anti-corrosive treatment to test the corrosion resistance of their products.suitable for parts, electronic components, metal materials, protective coatings and industrial products.

1, Durable hard 8 mm PVC construction, anti-corrosion and easy to clean.

2, Reservoirs with mixing system on stand.

3, Separate heating system apart from salt testing chamber to ensure accurate result and protect circuit.

4, Exposure zone lined with corrosion-resistant PVC sheets.

5, Clear cover allows viewing of test conditions.

6, Peaked cover prevents dripping of condensation onto samples.

7, Water seal and water level protection system.

8, Heated humidifying tower with air relief valve, adjustable tower tip to orient salt fog.

9, Automatic DI water fills to humidifying tower.

10, Digital temperature control.

11, Integral over-protection on heaters.

12, Low water cut-off protection on humidifying tower.


Method of operation:

1, the salt spray chamber internal parts: tower spray device, funnel, rod, V-type sample holder installed.

2, connected to the gas supply, the power is turned on, turn on the power switch.

3, in the pressure tank and cabinets add water, two low level light exterminate stop adding water, at the water seal groove seal.

4, purified water and sodium chloride salt in a ratio of 1:20 with a good. The solution was poured into the reagent with a good entrance (low saline to exterminate a lamp).

5, we can open all the lamps burn out operation, spray, time switch, switch demisting before opening end of the test.

6, according to the test requirements for equipment and cycles for continuous spraying mode (continuous spray T1, T2 tables are all set to 0, T1 is set to intermittent spray working time, T2 is time to rest).

Model

DR-H304-60

Internal Size (W*D*H)

600*450*400mm

External Size (W*D*H)

1060*580*1050mm

Chamber material

P.V.C rigid plastic plate, thickness 8mm

Sample tray

Diameter 10mm glass fiber rod, anti-rust

V-shape plastic steel to make sure sample in 15~30 degree inclined

Temperature Range

Room temperature ~ 55℃

Temperature Fluctuations

≤±0.5℃

Temperature Uniformity

≤±2℃

Temperature Precision

±1℃

Spraying method

Continuous and Periodic spray type

Test Chamber Temperature

Salt Spray Method (NSS ACSS)35±1℃
Corrosion-resistant Testing Method (CASS) 50±1℃

Saturated air Barrel Temperature

Salt Spray Method (NSS ACSS)47±1℃
Corrosion-resistant Testing Method (CASS) 63±1℃

Brine Temperature

35℃±1℃

Spray Quantity

1.0~2.0 ml / 80cm2 / hr

PH value

Salt Spray Method (NSS ACSS6.5~7.2)
Corrosion-resistant Testing Method(CASS)3.0~3.2

Lab Volume

108L

Brine Tank Capacity

25L

Multiple Safety Protection Devices

Current discharge protection, over pressure protection, over temperature protection, over load fuse protection

Accessories

Testing Salt/ Measuring tank/Nozzle Equipment x 1 set

Air source

1HP Air pump (provided by buyer)









Processed in 0.124413 Second , 126 querys.